- Temperature Instruments[5]
- Pressure Gauges[10]
- Other Measuring & Gauging Tools[10]
- Pumps[1]
- Gauges[2]
- Contact Person : Ms. Xu Xujiong
- Company Name : Yuyao City Mingtai Instrument Co., Ltd.
- Tel : 86-574-62802481
- Fax : 86-574-62802007
- Address : Zhejiang,Yuyao,No. 658, Fengshan Road, Qunli Village
- Country/Region : China
- Zip : 315400
SS Glycerine filled Pressure Gauge
1. Size Dial
40--80MM,100 mm (4”) & 150mm (6") White Anodised Aluminium marked in black finish
Single or dual scale
2. Mounting
Direct, Surface and Flush Pointer
Stainless steel coloured black
3. Case & Bezel Options: Micrometer Adjustable
Rugged one-piece full Safety Pattern case to S3 standard
with a solid baffle partition wall and full blow out back
4. Movement
manufactured in 1.6mm thick, 316L stainless steel. Stainless Steel Construction
Option : Viscous Damped movement to overcome the effects
5. Scale Ranges of minor pressure pulsations
0 to 600 mbar to 0 to 1400 bar Pressure
Equivalent units of pressure / vacuum available Window
Single and dual scales are available 3mm Laminated Safety Glass (Standard)
Option: Acrylic Plastic Window
6. Pressure Element
≤80 Bar – 316L Stainless Steel Bourdon Tube Environmental
7. Rating
≥81 Bar - 316L Stainless Steel Coil IP67 as defined in EN 60 529
8. Overload Traceability
Units withstands overload pressure up to 130% of FSD All instruments are individually calibrated and have an unique
Overload & vacuum stops are fitted on the movement Serial Number printed on the dial. A Certificate of Conformity
Traceable to National Standards is Supplied Free of Charge
Option: Mechanical overload clamps fitted internally to
enable units to withstand up to 3x the max scale reading
9. Certification available
BS EN 10204 3.1B Material Certification
10. Pressure Connection in 316 Stainless Steel Point by Point Test Certificate
3/8”, 1/2” BSP
1/2” NPT Safety
Other connections available, contact our Sales Dept for details All units are manufactured to comply with EN 837-1, S3
specification and other regulatory standards including PED.
11. Accuracy Class
CL:1 1.0% of FSD as defined in EN837-1 Installation
12. instructions
Option: 0.5% of FSD as defined in EN837-1 Refer to EN 837-2
13. Temperature Temperature Effect
Operating: -20 to +90 ºC Variation in indication caused by temperature shall not exceed
Storage: -40 up to +100 ºC ± 0.04 x (t2 – t1)% of the span where:
SS Glycerine filled Pressure Gauge